Measurement
SCPI.CALCulate(Ch).PARameter(Tr).DEFine
Attribute | Detail |
---|---|
Object Type | Property (read/write) |
Data Type | String |
Target | Trace Tr of channel Ch, Ch: channel number 1–16 (see Table 1 on page 25) Tr: trace number 1–16 (see Table 3 on page 33) |
Description | The measurement parameter of the trace. |
Allowable Values | "S11" : S11 parameter "S21" : S21 parameter "S12" : S12 parameter "S22" : S22 parameter "A" : Receiver A "B" : Receiver B "R1" : Receiver R1 "R2" : Receiver R2 |
Out of Range | An error occurs. Error code: 208. |
Preset Value | Depends on the trace number. Tr 1, Tr 5, Tr 9, Tr 13: "S11" Tr 2, Tr 6, Tr 10, Tr 14: "S21" Tr 3, Tr 7, Tr 11, Tr 15: "S12" Tr 4, Tr 8, Tr 12, Tr 16: "S22" |
Syntax | Dim Meas As String Meas = app.SCPI.CALCulate(Ch).PARameter(Tr).DEFine app.SCPI.CALCulate(Ch).PARameter(Tr).DEFine = "S11" |
Equivalent Softkeys | Measurement > Measurement > S11 / S21 / S12 / S22 Measurement > Measurement > Absolute > { A(1) / B(1) / R1(1) / A(2) / B(2) / R2(2) } |
Table 3. Tr: Trace Number | |
Data Type | Long |
Description | Trace number |
Range | from 1 to 16 |
Out of Range | An error occurs. Error code: 202. |
Notes | If the trace number is not specified, it is taken as equal to 1. |
SCPI.CALCulate(Ch).PARameter(Tr).SPORt
Attribute | Detail |
---|---|
Object Type | Property (read/write) |
Data Type | Long |
Target | Trace Tr of channel Ch, Ch: channel number 1–16 (see Table 1 on page 25) Tr: trace number 1–16 (see Table 3 on page 33) |
Description | The number of the source port for absolute measurements. |
Range | from 1 to 2 |
Out of Range | An error occurs. Error code: 208. |
Preset Value | 1 |
Syntax | Dim StimPort As Long StimPort = app.SCPI.CALCulate(Ch).PARameter(Tr).SPORt app.SCPI.CALCulate(Ch).PARameter(Tr).SPORt = 1 |
Equivalent Softkeys | Measurement > Measurement > Absolute > A(1) / B(1) / R1(1) / A(2) / B(2) / R2(2) |